Parametric cricothyrotomy trainer
Surgical cricothyrotomy · CC BY-NC-SA 4.0 · Original design for this repository
Half-shell larynx and trachea mounted on a strap plate. Wear it on a forearm or on the neck of a role-player, or strap it to a table edge. The cricothyroid membrane position is an open window in the print; the consumable skin layer over it is what gets cut.

Landmarks built in
- Thyroid cartilage with a superior V-notch (laryngeal prominence)
- Open cricothyroid membrane window (8 mm) between thyroid and cricoid
- Cricoid ring printed slightly proud so it palpates as a distinct step
- Four tracheal rings below, with an open 16 mm lumen for passing a tube or bougie
PLA or PETG, 0.2 mm layers, 3 or more perimeters, no supports; the model prints flat. PETG survives scalpels longer.
Assemble and run
- Thread 25 mm webbing or velcro through the four slots.
- Build the consumable skin: two or three layers of cloth tape, or Press'n Seal over a 3-5 mm foam sheet, stretched across the membrane window and the structures beside it. The trainee should have to find the membrane by feel, not by sight.
- Replace the skin layer between trainees. The print itself survives many sessions.
Technique sequence is the instructor's call; this model gives the hands realistic landmarks and a lumen that confirms correct placement.
Remix
cric_trainer.scad is fully parametric: lumen diameter, ring sizes, membrane gap, and plate dimensions are named variables at the top of the file. A pediatric or small-adult variant is a two-line edit.